Hukseflux FHF02SC Heat Flux Sensor

A next-level sensor from the world market leader in heat flux measurement, FHF02SC is a combination of our model FHF02 heat flux sensor and a heater. The heater allows the user to perform self-tests, verifying sensor functionality and stability, during use, without having to remove the sensor. FHF02SC is ideal for high-accuracy and long-term heat flux measurement, construction of calorimeters, (zero heat flux) core temperature measurement and thermal conductivity test equipment.

Suggested use

  • high-accuracy scientific measurement of heat flux, with a high level of data quality assurance
  • study of convective heat transfer mechanisms
  • calorimeter prototyping
  • (zero heat flux) non-invasive core temperature measurement
  • thermal conductivity test equipment

A Next-Level Sensor for General-Purpose Heat Flux Measurement

FHF02SC is a sensor for general-purpose heat flux measurement, combined with a heater. It is used when the highest level of quality assurance is required and for special experiments. The FHF02SC thermopile sensor measures heat flux (W/m²) through the object in which it is incorporated or on which it is mounted. Additionally the integrated type T thermocouple allows for temperature measurements. Both the thermopile and thermocouple are passive sensors; they do not require power. The combination of a heat flux sensor with a heater allows the user not only to measure a heat flux, but also to apply a heat flux.

Using FHF02SC is easy

Using FHF02SC’s heat flux sensor is easy. It can be connected directly to commonly used data logging systems. The heat flux sensor output is a voltage signal that is proportional to the heat flux through the sensor: the heat flux in W/m² is calculated by dividing the sensor output, a small voltage, by the sensitivity. The sensitivity is provided with FHF02SC on its product certificate.

Measurand: Heat flux, Temperature
Measurement range: (-10 to +10) x 10³ W/m²
Sensitivity (nominal): 5.5 x 10⁻⁶ V/(W/m²)
Temperature sensor: type T thermocouple
Thermal spreaders: included
On-line functionality testing: self-test including self-calibration
Rated bending radius: infinite, bending is not recommended
Sensing area: 9 x 10⁻⁴ m²
Sensor thermal resistance: 37 x 10⁻⁴ K/(W/m²)
Sensor resistance range: 50 to 100 Ω
Sensor thickness: 0.9 x 10⁻³ m
Uncertainty of calibration: ±5% (k = 2)
Operating temperature range: -40 to +150°C
IP protection class: IP67
Standard wire length: 2 m
Heater resistance: 100 Ω (nominal)
Heater power supply: 12 VDC
Options: longer wire length upon request
  • Hukseflux HTR01 Heater
    Hukseflux, the world market leader in heat flux sensors, simplifies heat flux sensor calibration. HTR01 is a heater with 4-wire connection with a known surface area and electrical resistance. It is used for calibration and functionality checks of FHF-type heat flux sensors. Users can now easily and objectively check their sensor performance before and after use. See also model FHF02SC heat flux sensor with integrated heater.