The AWQ-NTU sensor is an optical sensor for the measurement of Nephelometric Turbidity.
• IR optical sensor with optical fibre
• Range: 0 to 4000 NTU or 0-4500 mg/L
• Robust and waterproof (IP68)
• Ultra low-power consumption
• Digital output: Modbus RS-485 or SDI-12
APPLICATIONS:
• Urban wastewater treatment (inlet/ outlet controls)
• Sanitation network
• Industrial effluent treatment
• Surface water monitoring
• Drinking water
Utilising the AWQ-NTU Sensor in a custom IoT (Internet of Things)
The SNiP-NTU is a ‘Sensor Node Integrated Package’ for LoRaWAN or CAT-M1 communication of real-time nephelometric turbidity and temperature measurements for continuous water quality monitoring.
The base SNiP-NTU integrates 1x AWQ-NTU Sensor with 1x S-NODE to a site’s unique network, communication and power requirements. The S-NODE can support an added 3x compatible sensors.
• The measurement principle is based on IR nephelometry/850 nm. The sensor can be calibrated with a formazine standard solution.
• The AWQ-NTU sensor uses low-cost optical technology, requiring very little maintenance and no consumables.
The sensor can be connected to any types of transmitters, display units, controllers, or data loggers with Modbus RS-485 or SDI-12 inputs. The optical sensor saves its calibration data for better measurement management.
All data concerning calibration, history, users and measurements are processed within the AWQ-NTU sensor and transmitted via RS-485 or SDI-12.
Compact, robust and light, the PVC sensor allows for handheld or fixed unit application.
Measurement: | |
Measurement Principle: | Diffusion IR at 90° |
Measure range: | 0 to 4000 NTU in 5 ranges: • 0 – 50 NTU • 0 – 200 NTU • 0 – 1000 NTU • 0 – 4000 NTU • AUTOMATIC OR 0 to 4500 mg/L |
Calibration: | Range 0-500 mg/L according to NF EN 872 Range >500 mg/L according to NF T 90 105 2 |
Resolution: | 0.01 to 1 NTU – mg/L |
Accuracy: | < ±5% |
Temperature Measurement: | NTC Thermistor |
Interface: | Modbus RS-485 (standard); SDI-12 (optional) |
Measurement Frequency: | 1Hz |
Power Requirement: | 5 to 12 volts DC |
Standby Current Draw: | 40 μA |
Average RS485 Draw (1 measurement/second): | 820 μA |
Average SDI12 Draw (1 measurement/second): | 4.2 mA |
Current Pulse: | 500 mA |
Sensor: | |
Operating Temperature: | 0°C to +50°C |
Storage Temperature: | -10°C to +60°C |
Dimensions: | Diameter: 27 mm; length: 170 mm |
Weight: | 300g (sensor and 3 meter cable) |
Construction: | PVC, DELRIN, Quartz, PMMA, Polyamide |
Maximum Pressure: | 5 bar |
Connection: | 9 armoured connectors, polyurethane jacket, bare-wires or waterproof Fischer connector |